Problem: 
How to set a power reference on the Viavi Powerchek OP-1 power meter?
Solution: 
To set a reference on the Powerchek connect the light source directly to the Powerchek. Set the desired wavelength on the light source and turn the laser on. Source should be set for CW.  Set the wavelength on the Powerchek by touching on the wavelength box until the correct wavelength shows. Set the scale selection for dBm by touching on that selection until it says dBm. Verify that you have a good dBm reading showing in the display. Press the capture button (large triangular hard button above the display) once. The power reading will freeze and show highlighted. Press and hold the REF selection until the unit beeps. It will switch automatically to the db scale and should read around 0 db. If there is no REF selection visible in the display touch in the bottom left corner to bring it up to the screen.

Product Family:

Wir stehen Ihnen immer zur Seite

Bei uns erhalten Sie den Support, die Services, die umfassenden Schulungen und die Ressourcen, die Sie benötigen. Diese Leistungen tragen zur Maximierung Ihrer VIAVI Investitionen bei.

Wertschöpfende Dienstleistungen, die Ihr Portfolio an Systemlösungen und Geräten von VIAVI abrunden, um möglichst niedrige Gesamtbetriebskosten sicherzustellen

Der Kundendienst stellt Ihnen RMA-Rücksendenummern für die Reparatur und Kalibrierung aus

Technische Schulungen, Produkttraining und integriertes Lernen für Techniker, die mit unseren neuen Produkten oder auch bereits vorhandenen Lösungen möglichst effizient arbeiten möchten

Die Technical Assistance Center (TAC) unterstützen Sie bei der Nutzung/Konfiguration der Produkte und helfen Ihnen auch bei Leistungsstörungen