Problem: 
How to set a power reference on the Viavi Powerchek OP-1 power meter?
Solution: 
To set a reference on the Powerchek connect the light source directly to the Powerchek. Set the desired wavelength on the light source and turn the laser on. Source should be set for CW.  Set the wavelength on the Powerchek by touching on the wavelength box until the correct wavelength shows. Set the scale selection for dBm by touching on that selection until it says dBm. Verify that you have a good dBm reading showing in the display. Press the capture button (large triangular hard button above the display) once. The power reading will freeze and show highlighted. Press and hold the REF selection until the unit beeps. It will switch automatically to the db scale and should read around 0 db. If there is no REF selection visible in the display touch in the bottom left corner to bring it up to the screen.

Product Family:

Asistencia en cada paso

Proporcionamos soporte técnico, servicios, una formación completa y los recursos que necesita. Todo ello forma parte de lo que hacemos con el fin de maximizar el valor de su inversión en VIAVI.

Valor y servicios que complementan su catálogo de instrumentos y soluciones de sistemas VIAVI para proporcionar un coste total de propiedad excelente

Ofrecemos un servicio de atención al cliente que proporciona solicitudes de autorización de devolución de material (RMA) para tareas de reparación y calibración

Soluciones educativas técnicas, formación sobre productos y una formación combinada para técnicos que utilizan productos nuevos o trabajan con herramientas existentes

El centro de asistencia técnica (TAC) le ayudará a utilizar y configurar los productos y se encargará de los problemas relativos al rendimiento del producto