Knowledge Base - Solution

TB2000 inspection probe test button not testing.

Mardi, Décembre 31, 2019
Soumis par ryan.grace le 8 Juillet 2021 - 1:47pm
Problem: 
When I press the test button on the Viavi P5000i inspection probe it does not provide the zone pass/fail information. I still just see an image and it appears to be frozen when I use the probe on my MTS/TB2000.
Solution: 
There is a SETUP parameter, on the MTS/TB2000, that controls the function of the test button on the P5000i. Go to SETUP and click on Analysis. Three is a field called capture button that can be set to freeze or freeze and test. To test it must be set for Freeze and Test. If you have no such setting under analysis then check the name on the body of the probe. It is most likely a P5000 and not a P5000i. The P5000 cannot test on its own and you must press TEST on the screen instead.

Product Family:

Suivi à chaque étape

Nous assurons l’assistance, les services, des formations complètes ainsi que les ressources qu’il vous faut. Tout cela dans le but de maximiser la valeur de vos investissements VIAVI.

Des prestations de service à valeur ajoutée qui optimisent le retour sur investissement de votre solution système et de votre gamme d’instruments VIAVI"

Un service client qui émet les autorisations de retour de matériel (Return Material Authorization, RMA) pour les réparations et la calibration.

Des solutions de formation technique, une formation à l’utilisation des produits et un enseignement mixte pour les techniciens qui utilisent de nouveaux produits ou travaillent avec les outils existants

Le centre d’assistance technique de VIAVI peut vous aider à utiliser/configurer vos produits ou à résoudre des problèmes liés aux performances des produits