Problem: 
How to set a power reference on the Viavi Powerchek OP-1 power meter?
Solution: 
To set a reference on the Powerchek connect the light source directly to the Powerchek. Set the desired wavelength on the light source and turn the laser on. Source should be set for CW.  Set the wavelength on the Powerchek by touching on the wavelength box until the correct wavelength shows. Set the scale selection for dBm by touching on that selection until it says dBm. Verify that you have a good dBm reading showing in the display. Press the capture button (large triangular hard button above the display) once. The power reading will freeze and show highlighted. Press and hold the REF selection until the unit beeps. It will switch automatically to the db scale and should read around 0 db. If there is no REF selection visible in the display touch in the bottom left corner to bring it up to the screen.

Product Family:

Suivi à chaque étape

Nous assurons l’assistance, les services, des formations complètes ainsi que les ressources qu’il vous faut. Tout cela dans le but de maximiser la valeur de vos investissements VIAVI.

Des prestations de service à valeur ajoutée qui optimisent le retour sur investissement de votre solution système et de votre gamme d’instruments VIAVI"

Un service client qui émet les autorisations de retour de matériel (Return Material Authorization, RMA) pour les réparations et la calibration.

Des solutions de formation technique, une formation à l’utilisation des produits et un enseignement mixte pour les techniciens qui utilisent de nouveaux produits ou travaillent avec les outils existants

Le centre d’assistance technique de VIAVI peut vous aider à utiliser/configurer vos produits ou à résoudre des problèmes liés aux performances des produits