40G Certifier Probe mismatch error on reference.

Monday, August 7, 2023
Submitted by ryan.grace on August 11, 2023 - 8:43pm
Problem: 
I am trying to perform a copper reference on my Viavi 40G Certifier and getting a probe mismatch error message.
Solution: 
You cannot perform a reference with 2 channel adapters. A reference needs to be done with a permanent link adapter on the local unit and a channel adapter on the remote using one permanent link cable.