40G Certifier Remote Calibration Data Missing

Monday, July 24, 2023
Submitted by ryan.grace on April 4, 2024 - 10:40am
Problem: 
The remote unit calibration date is intermittently missing from report data for some test results on my 40G Certifier.
Solution: 
The most likely cause of the calibration date being missing in the report data is procedural. The SAVE must be performed and completed prior to breaking the physical path between the Local and the Remote unit. If a cable or fiber is disconnected prior to the save completing then the calibration date of the remote unit will be missing. You should wait for the SAVE to complete before moving any cables/fibers from either the Local or the Remote units.