Knowledge Base - Solution

HST Intermittent issues with FED

Monday, July 31, 2023
Submitted by ryan.grace on April 4, 2024 - 6:37am
Problem: 
I am having intermittent issues operating my HST or ONX with the FED (Far End Device). Sometimes it cannot communicate or I get inconsistent readings.
Solution: 
1. Make sure the 8 pin mod connector and jack are clean and fully seated into the jack on the FED. 2. Make sure the battery on the FED is not low. Replace battery if in question. 3. Clip the unsued clips on the FED cable to the back strap of the FED glove (soft case) to make sure that no usused clips can come into contact with anything. Make sure clips are not touching each other. Do not hang the FED by connector leads. Use the D Ring on the FED glove (soft case) to support the FEDs weight if necessary. 4. Send an Open All command to the FED first before sending the test command. 5. If performing resistance readings be sure to use leakage mode if normal mode shows an unexpected reading. 6. Make sure the copper leads on the HST are fully reseated into their jacks on the unit and try replacing the leads. You can use any color lead in any jack. Perform a lead compensation.

Product Family:

Support at Every Step

We provide support, services, comprehensive training and the resources you need. It’s all part of what we do to maximize the value of your VIAVI investment.

Value add services that compliment your VIAVI system solution and instrument portfolio to provide a total cost of ownership

Customer Service that provides Return Material Authorization (RMA) for repair and calibration

Technical education solutions, product training, and blended learning for technicians who are using new products or working with existing tools

Technical Assistance Center to assist you in using/configuring products or address issues regarding product performance